UDI Medical Devices: Types Explained (EU MDR Guide)

UDI medical devices are not just a labelling requirement. They are the backbone of traceability under EU MDR.
UDI-Medical-Devices-Types-Explained-EU-MDR-Guide

Updated: 21st May 2026

Reviewed by: Steve Condie, MTOPRA (Regulatory Manager)

Understanding Basic UDI-DI, UDI-DI and UDI-PI Under EU MDR

UDI means Unique Device Identifier, it is implemented within latest regulations to assist with medical device and IVD (In Vitro Diagnostic) device traceability. In this blog we discuss this complex topic.

Annex VI, Part C

01. Reusable Devices

Devices that require cleaning and sterilization between patients have the strictest physical requirements under EU MDR.

  • Permanent Marking: UDI must be on the device itself (Direct Part Marking).
  • Durability: Must remain legible throughout the entire expected lifetime of the device.
  • Reprocessing: The UDI carrier must withstand multiple cycles of disinfection and sterilization.
  • Traceability: Linked to the specific reprocessing cycle records.
Annex VI, 6.5

02. Software (SaMD)

Medical software does not have physical packaging in the traditional sense, so the UDI is digital.

  • Internal Display: UDI must be accessible via a menu (e.g., "About" or "System Info" screen).
  • Version Control: Every major version change requires a new UDI-DI.
  • PI Requirements: The UDI-PI must include the software version number rather than a batch code.
  • Distribution: UDI must be included on any physical media (USB/CD) or download site.
Article 22

03. Systems & Packs

When multiple devices are combined into a single kit, the UDI requirements apply to both the pack and the contents.

  • The Pack UDI: The system itself must have its own unique Basic UDI-DI and UDI-DI.
  • Individual Components: Each device within the pack must have its own UDI unless they are not CE marked separately.
  • EUDAMED: The "System/Procedure Pack Producer" must register the kit under their own Actor ID.
  • Sterilization: If the pack is sterilized, a new UDI-PI must be assigned for that process.
Class III

04. Implantable Devices

As the highest risk class, these devices require the most comprehensive digital "paper trail."

  • Implant Cards: UDI must be present on the mandatory Implant Card provided to the patient.
  • Storage: Healthcare institutions must store the UDI of implantable devices they have received.
  • SSC: The UDI-DI must be linked to the Summary of Safety and Clinical Performance.
  • Rapid Recalls: Enables targeted tracking if an individual batch shows performance issues.

⚡ Quick Summary

  • Basic UDI-DI: For administration/EUDAMED (not on label).
  • UDI-DI: Identifies the specific model (on label).
  • UDI-PI: Identifies the batch/expiry (on label).
UDI Numbers explained Medical Devices and IVDs under EU MDR and IVDR

What is UDI in medical devices?

Compliance

Accurate Identification

UDI eliminates the ambiguity of similar product names by providing a globally unique numeric key for every version and variant of a device.

Safety

Efficient Recall Management

In the event of a safety alert, manufacturers can pinpoint affected batches instantly, preventing unnecessary disruption to clinical operations.

Surveillance

PMS Data Integration

Post-Market Surveillance relies on clean data. UDI allows for the seamless aggregation of incident reports and clinical performance metrics.

Efficiency

Supply Chain Integrity

Distributors and importers can automate inventory tracking, reducing manual entry errors and ensuring stock rotation is compliant with expiry dates.

Legal

EUDAMED Synchronicity

The UDI acts as the bridge between your physical product and the EUDAMED database, ensuring the public and regulators see consistent data.

Global

Market Access Readiness

By following EU MDR UDI standards, you align your technical file with IMDRF guidelines, significantly lowering the barrier for entry into non-EU markets.

UDI requirements under EU MDR

The UDI Requirements EU MDR are operational, not theoretical.

Manufacturer Obligations

MDCG 2021-19: Makes it clear that Basic UDI-DI must be assigned before placing a device on the market and before Notified Body submission.

Strategic Integration

UDI is not an isolated task. It is the connective tissue linking your core regulatory assets:

If your UDI structure is weak, your entire compliance framework is unstable.

UDI hierarchy explained

This is where things either click or fall apart. UDI is not a single identifier—it is a structured hierarchy. Understanding this is the difference between compliance and chaos.

Device Family Level

Basic UDI-DI

The anchor of your UDI system. It identifies a group of devices at a model or family level.

  • Used in Certificates & SSCP
  • Declaration of Conformity
  • Technical Documentation
  • Note: Does NOT appear on labels
Commission View: Described as the main key for linking regulatory documentation and database records.
Specific Device Level

UDI-DI

Identifies a specific device version or configuration. It is fixed and assigned per packaging level.

  • Mandatory on the Label
  • Required for Sterile vs Non-Sterile
  • Distinct for each packaging level
MDCG Warning: UDI-DIs must be assigned before market placement. Many companies realize too late their SKU structure is non-compliant.
Production Level

UDI-PI (Production Identifier)

The variable element that makes real-world traceability possible across the supply chain.

  • Lot / Batch Numbers
  • Serial Numbers
  • Expiry & Manufacturing Dates
Unlike the UDI-DI, the PI is dynamic and changes with every production run.

Special Exception: Master UDI-DI

While not relevant for most medical devices, the Master UDI-DI applies specifically to highly customized categories like contact lenses, spectacle frames, and ready-to-wear reading glasses.

UDI hierarchy summary

Identifier What it identifies On Label EUDAMED Documentation Type
Basic UDI-DI Device family/group No Yes Yes Fixed
UDI-DI Specific device/version Yes Yes Yes Fixed
UDI-PI Production-level data Yes Traceability only No Variable
Master UDI-DI Special device grouping No Yes Yes Fixed

How UDI works in practice

This is where theory meets reality.

A Compliant UDI System Requires:

  • Correct identifier assignment
  • Proper label and packaging integration
  • Structured data for EUDAMED

Manufacturers Must:

  • Place the UDI carrier on labels or packaging
  • Ensure identifiers match documentation
  • Register device data in EUDAMED
"The European Commission confirms that manufacturers are responsible for UDI carrier placement and data accuracy."

Critical Milestone: 28 May 2026

From this date, the UDI/Device registration module in EUDAMED becomes mandatory. This turns UDI from a regulatory requirement into a critical operational dependency.

Medical device UDI numbers

UDI-DI Triggers: When Does a Change Require a New ID?

A frequent point of failure in Post-Market Surveillance is the "silent change"—making a modification to a device without updating its UDI-DI. According to EU MDR Annex VI, Part C, any change that could lead to the misidentification of the device or its traceability requires a new UDI-DI.

Brand or Model Name A change to the trade name or the specific model version. If the user perceives it as a different product, the regulator requires a different ID.
Risk Classification If a device is up-classified (e.g., from Class IIa to IIb), the UDI-DI must be replaced to reflect the new regulatory status in EUDAMED.
Quantity in a Package Moving from a single-pack to a 10-pack is not just a commercial change; it is a packaging-level UDI-DI trigger.
Critical Clinical Attributes Changes to "Sterile" vs "Non-Sterile" status or "Single Use" vs "Reusable" designations are mandatory triggers for a new UDI-DI.
The "Significant Change" Rule: For legacy devices under Article 120, a UDI-DI trigger often coincides with a "significant change" that would invalidate your legacy status, forcing an immediate transition to full EU MDR compliance.

Beyond the Manufacturer: Economic Operator Duties

UDI is not solely the manufacturer's burden. Article 25 of the MDR places strict verification duties on Importers and Distributors. If you are an Economic Operator, you are legally required to verify that the UDI is present and correctly registered.

Verification

Importers must verify that the UDI has been assigned and the carrier is affixed before placing the device on the market.

Sample Checking

Distributors must perform representative sampling to ensure that the UDI data on the label matches the documentation.

Storage Records

All Economic Operators must store and keep, preferably by electronic means, the UDI of Class III implantable devices.

Failure to verify UDI data makes an Importer or Distributor liable for the non-compliance. This creates a "network effect" where your customers (distributors) will eventually reject shipments if your UDI data structure is flawed.

Common UDI Mistakes

These show up again and again in manufacturer audits. They are rarely minor—they create systemic failures.

Labeling Errors Treating Basic UDI-DI as a label identifier.
Timing Failures Assigning UDI-DIs too late in the process.
Weak Grouping Flawed product grouping logic that fails audits.
Packaging Oversight Ignoring unit, shelf, or case-level identifiers.
Data Unreadiness Approaching EUDAMED as a simple "data entry" task.
The SKU Fallacy Assuming internal SKU structures equal UDI compliance.
These are not minor issues. They create downstream failures across documentation, audits, and registration that can stall market access.

How to Implement a UDI System

A practical, scalable implementation approach for high-stakes compliance.

1
Group & Assign Define product groupings and assign Basic UDI-DIs.
2
Device Identification Assign UDI-DIs for all configurations and packaging levels.
3
Production Tracking Define the UDI-PI structure for real-world traceability.
4
Process Alignment Align labelling and artwork processes with digital records.
5
Registration Readiness Prepare and validate data for EUDAMED registration.
6
Lifecycle Control Build change control rules to ensure global scalability.
Global Standards: GS1 explains that UDI supports globally harmonised identification. In GS1 terms, the Basic UDI-DI is represented by the Global Model Number (GMN).

UKCA and Future UK UDI Strategy

Let’s keep it grounded: UK requirements are still evolving, but market access depends on clarity.

EU MDR Scope
Applies in EU Member States and Northern Ireland.
Great Britain (UK MDR)
Operates under UK MDR 2002. Non-UK manufacturers must appoint a UKRP.

The MHRA confirms that devices placed on the GB market must be registered. The key takeaway: Do not build a rigid, EU-only UDI system. Build something scalable.

BUILD A SCALABLE, FUTURE-PROOF UDI SYSTEM

How Patient Guard supports UDI compliance

UDI Strategy and Scoping

  • Portfolio Analysis: We audit your product range to identify UDI requirements and implementation timelines.
  • Risk Class Assessment: Determining Class I, IIa, IIb, or III status to dictate EUDAMED deadlines.
  • Legacy Device Strategy: Navigating the transition for MDD-certified devices into the UDI system.

Basic UDI-DI & Structuring

  • The Regulatory Umbrella: Defining the Basic UDI-DI as the primary key for your EUDAMED registration.
  • Granular Traceability: Assigning UDI-DIs for every packaging level (unit, box, carton).
  • Issuing Entity Guidance: Expert advice on selecting and managing codes via GS1, HIBCC, or IFA.

EUDAMED Readiness

  • Actor Registration: Ensuring correct registration for Manufacturers, Reps, and Importers.
  • Data Attribute Mapping: Mapping internal data to the 100+ mandatory EUDAMED attributes.
  • Validation & Submission: "Pre-flight" data checks to ensure first-time submission success.

Labelling Alignment

  • AIDC & HRI Integration: Ensuring labels meet both barcode and human-readable requirements.
  • Symbol Implementation: Integrating ISO symbols for sterile status, expiry, and batch numbers.
  • Direct Part Marking (DPM): Guidance on permanent marking for reusable medical devices.

Notified Body Preparation

  • Technical Documentation: Updating Technical Files to include robust UDI traceability data.
  • QMS Integration: Embedding UDI management into your Quality Management System and SOPs.
  • Audit Defense: Preparing your team for specific NB questions regarding assignment logic.

EU MDR Compliance Integration

  • Post-Market Surveillance: Leveraging UDI for faster tracking of field safety actions (FSCA).
  • Clinical Evaluation: Linking clinical data to specific identifiers for accurate CER reports.
  • Supply Chain Transparency: Ensuring Economic Operators verify UDI data at every stage.

The Golden Thread: Physical Labeling vs. Digital Records

Under EU MDR, the UDI is often described as the "Golden Thread" that connects a physical medical device in a hospital ward to its regulatory history in EUDAMED. This connection is not merely a label; it is a live data link. Manufacturers often underestimate the complexity of ensuring that the Human Readable Interpretation (HRI) and the Automatic Identification and Data Capture (AIDC)—typically a 2D DataMatrix or Barcode—are perfectly synchronized.

If the barcode scans a UDI-DI that does not match the printed text, the device is technically mislabeled and non-compliant. Furthermore, this data must be mirrored exactly in the EUDAMED UDI/Device module. Any discrepancy between the physical label, the Technical File, and the EUDAMED entry is a "major non-conformity" during a Notified Body audit. Maintaining this 1:1:1 ratio across thousands of SKUs is where operational excellence is tested.

The Legacy Device Challenge (Article 120)

A significant point of confusion remains for "Legacy Devices"—those placed on the market under MDD or AIMDD certificates during the transition period. While Article 120 of the EU MDR provides an extension for certificate validity, it does not exempt these devices from UDI requirements.

Legacy devices must still be assigned a Basic UDI-DI and UDI-DI for registration in EUDAMED, even if the physical label does not yet carry the UDI carrier. However, the moment a "significant change" in design or intended purpose occurs, the device loses its legacy status and must fully comply with MDR labeling. Navigating this "grey zone" requires a documented strategy to avoid premature label changes or, conversely, missing the deadline for digital registration.

What Notified Bodies Look For in a UDI Audit

During a Quality Management System (QMS) audit under ISO 13485:2016 and EU MDR, auditors will move beyond checking if a barcode exists. They will perform a vertical trace. They will select a random device from your warehouse and ask to see the link between the UDI-PI on that box and the production batch record, the sterilization validation, and the EUDAMED registration.

Auditors expect to see a formal UDI Procedure within your QMS. This procedure must define who is responsible for data accuracy, how UDI-DIs are issued for new products, and how "triggers" for a new UDI-DI are identified (e.g., changes to the brand name, device model, or clinical use). If your QMS does not explicitly govern UDI lifecycle management, you are at risk of a finding.

UDI as an Operational Dependency

While UDI is a regulatory burden, the most sophisticated manufacturers view it as a supply chain upgrade. A robust UDI system enables High-Precision Traceability. In the event of a Field Safety Corrective Action (FSCA) or a recall, a functional UDI-PI system allows you to pinpoint exactly which batches are affected, rather than performing a "blanket recall" that destroys profit and brand reputation.

By 2026, hospitals and procurement bodies across Europe will rely on UDI for inventory management. If your UDI data is structured poorly, your products may be "invisible" to automated hospital systems, creating an unofficial barrier to sale. UDI is no longer just about compliance; it is about your ability to transact in a digital healthcare economy.

The Final Verdict

UDI for medical devices is not about generating identifiers. It is about building regulatory infrastructure.

Basic UDI-DI
UDI-DI
UDI-PI
Master UDI-DI

Done Properly

UDI becomes a strategic asset supporting:

  • End-to-end traceability
  • Global regulatory consistency
  • Permanent audit readiness
  • Multi-market scalability

Done Poorly

UDI becomes a liability that creates:

  • Compliance process friction
  • Documentation inconsistencies
  • Registration delays
  • Increased audit risk

The foundation you build today determines your market access tomorrow.

Watch our YouTube Video about UDI

Frequently Asked Questions About UDI Types

No. Unlike the UDI-DI, the Basic UDI-DI is a purely administrative identifier used for regulatory documentation (like the Declaration of Conformity) and EUDAMED registration. It does not appear on the physical label or the packaging.

You must assign a new UDI-DI whenever there is a change that could lead to the misidentification of the device or its traceability. This includes changes to the device name/model, its risk class, its "single-use" status, or its sterile packaging.

Yes. Reusable devices that require cleaning, disinfection, or sterilization between uses must have a permanent UDI carrier (direct marking) on the device itself. This ensures traceability throughout the product's entire lifecycle, even if the original packaging is long gone.

UDI-DI (Device Identifier): The static part of the code that identifies the specific model and manufacturer.

 

UDI-PI (Production Identifier): The dynamic part of the code that includes production-specific data like the batch number, serial number, expiry date, or manufacturing date.

Yes. Software that is commercially available on its own and qualifies as a medical device must have a UDI. For software, the UDI is typically assigned at the system level and should be displayed on a "header" or "about" screen within the user interface.

Under the current EU MDR timeline, the UDI/Device registration module is expected to become mandatory by May 2026. However, manufacturers are encouraged to register early to ensure compliance and smooth market access.

References

This guide is based on the following legislation and official regulatory guidance relating to the Unique Device Identification (UDI) system under Regulation (EU) 2017/745 (MDR).

Organisation Reference Why it's relevant
European Union Regulation (EU) 2017/745 on Medical Devices (MDR) Establishes the legal framework for the Unique Device Identification (UDI) system, including the requirements in Chapter III and Annex VI governing Basic UDI-DI, UDI-DI, UDI-PI, device registration and traceability.
European Commission Unique Device Identifier (UDI) Provides the European Commission's official overview of the UDI system, explaining its objectives, structure and role in improving medical device traceability and patient safety.
European Commission UDI/Device Registration in EUDAMED Explains how manufacturers register UDI and device information within EUDAMED, including the requirements for device registration, UDI datasets and ongoing regulatory obligations.
Medical Device Coordination Group (MDCG) MDCG 2018-1 Rev.4 – Guidance on Basic UDI-DI and Changes to UDI-DI Provides official guidance on Basic UDI-DI, UDI-DI assignment, grouping principles and the circumstances that require a new UDI-DI to be assigned.
European Commission UDI Helpdesk Rules for Specific Device Types Explains how UDI requirements apply to specific device categories, including reusable devices, software, configurable devices and systems or procedure packs, supporting the practical examples discussed throughout this guide.

UDI requirements, EUDAMED functionality and regulatory guidance continue to evolve. Manufacturers should always consult the latest published legislation and official guidance when implementing, maintaining and updating compliant Unique Device Identification systems under the MDR.

Steve Condie, MTROPRA

Steve Condie, MTROPRA

Reviewed by
Steve Condie, MTOPRA
Regulatory Manager | ISO 13485 Internal Auditor
20+ years in medical device regulatory affairs, ISO 10993 biocompatibility, MDR/IVDR compliance and quality systems.

Patient Guards Recent Posts

Cosmetic Product Safety Report (CPSR): A Complete Guide to UK Cosmetic Compliance

Before a cosmetic product can legally be placed on the UK market, manufacturers and Responsible Persons must demonstrate that it is safe for human use under normal or reasonably foreseeable conditions. The Cosmetic Product Safety Report (CPSR) is one of the most important regulatory documents required under the UK Cosmetics Regulation. This guide explains what a CPSR is, who can prepare one, what information it must contain, how it relates to the Product Information File (PIF) and how it supports legal cosmetic compliance.

Read More »

IVDR PMPF Explained: A Complete Guide to Post-Market Performance Follow-up

Post-Market Performance Follow-up (PMPF) is a fundamental requirement under the EU In Vitro Diagnostic Regulation (IVDR), ensuring that manufacturers continually monitor the scientific validity, analytical performance and clinical performance of their in vitro diagnostic medical devices after CE marking. This guide explains IVDR PMPF requirements, PMPF Plans, PMPF Reports, Annex XIII expectations and how ongoing performance monitoring supports continued regulatory compliance throughout the device lifecycle.

Read More »

IVDR Scientific Validity Explained: A Complete Guide for Manufacturers

Scientific Validity is the first pillar of IVDR Performance Evaluation and provides the scientific foundation demonstrating that an analyte or biomarker is associated with a specific clinical condition or physiological state. This guide explains Scientific Validity under Regulation (EU) 2017/746, including literature reviews, Scientific Validity Reports, Annex XIII requirements, evidence appraisal and how Scientific Validity supports successful CE marking.

Read More »
Share this guide:
Posted on Google Google
Munna P profile picture
Munna P
52 days ago
Google star 1Google star 2Google star 3Google star 4Google star 5Trustindex verifies that the original source of the review is Google.
Working with the Patient Guard team has been a great experience throughout our MHRA and ISO 13485 documentation journey. Their expertise, structured approach, and practical guidance helped our team build a robust quality management system while keeping us aligned with regulatory expectations. The collaboration was professional, responsive, and focused on finding solutions rather than simply identifying issues. A special thank you to Alex and Steve for their outstanding coordination, responsiveness, and continuous support throughout the project. They were always approachable, provided valuable feedback, and worked closely with our team to resolve challenges efficiently. Their commitment made a significant difference in keeping our documentation effort on track. I highly recommend Patient Guard to any healthcare or MedTech organization looking for experienced regulatory and quality system partners for MHRA, ISO 13485, and broader medical device compliance initiatives. Thank you again to the entire Patient Guard team for being such reliable partners.
Posted on Google Google
Peter Reeve profile picture
Peter Reeve
79 days ago
Google star 1Google star 2Google star 3Google star 4Google star 5Trustindex verifies that the original source of the review is Google.
STEPPER design, manufacture & distribute eyewear across the globe. With the increasingly complex landscape concerning the placing of Mecial Devices onto the market, we realised we needed professional guidance. We found Patient Guard via a simple internet search and are delighted we did! They provide a pragmatic solution to our needs, are totally reliable & always available to answer our (often simplistic) questions. They are highly efficient & responsive to what is a changing picture in our world and nothing is too much trouble. We have a much better understanding of regulatory affairs and our responsibilities as manufacturers & distributors and they support us in navigating the requirements in different territories. Updating our Declaration of Conformity, ensuring our labelling is compliant and acting as our PRRC are the key areas of their service for us.
Posted on Google Google
Derek Timm profile picture
Derek Timm
79 days ago
Google star 1Google star 2Google star 3Google star 4Google star 5Trustindex verifies that the original source of the review is Google.
For those companıes lookıng to comply to ISO standards and ın partıcular ISO13485 whıch to be honest ıs a nıghtmare I would strongly suggest goıng to the professıonals as ındeed we dıd by joınıng forces wıth Patıent Guard Ltd The staff are fantastıc nothıng ıs too much trouble and as a medıcal supply company we sımply cannot lıve wıthout them Thanks ın partıcular to Alex and Steve for all the hard work and our best regards from Dan Medıca South Lımıted
Posted on Google Google
BMSCriticalCare profile picture
BMSCriticalCare
116 days ago
Google star 1Google star 2Google star 3Google star 4Google star 5Trustindex verifies that the original source of the review is Google.
Great service, very helpful and always willing to answer any questions we have,
Posted on Google Google
Thomson Software profile picture
Thomson Software
787 days ago
Google star 1Google star 2Google star 3Google star 4Google star 5Trustindex verifies that the original source of the review is Google.
Alex Lewis of PatientGuard guided us through the ISO13485 process in a thorough, systematic and efficient manner. He was friendly, patient and willing to go the extra mile. Excellent service.
Verified by Trustindex
Trustindex verified badge is the Universal Symbol of Trust. Only the greatest companies can get the verified badge who has a review score above 4.5, based on customer reviews over the past 12 months. Read more

Most Popular

Cosmetic Product Safety Report (CPSR): A Complete Guide to UK Cosmetic Compliance

Before a cosmetic product can legally be placed on the UK market, manufacturers and Responsible Persons must demonstrate that it is safe for human use under normal or reasonably foreseeable conditions. The Cosmetic Product Safety Report (CPSR) is one of the most important regulatory documents required under the UK Cosmetics Regulation. This guide explains what a CPSR is, who can prepare one, what information it must contain, how it relates to the Product Information File (PIF) and how it supports legal cosmetic compliance.

Read More »

IVDR PMPF Explained: A Complete Guide to Post-Market Performance Follow-up

Post-Market Performance Follow-up (PMPF) is a fundamental requirement under the EU In Vitro Diagnostic Regulation (IVDR), ensuring that manufacturers continually monitor the scientific validity, analytical performance and clinical performance of their in vitro diagnostic medical devices after CE marking. This guide explains IVDR PMPF requirements, PMPF Plans, PMPF Reports, Annex XIII expectations and how ongoing performance monitoring supports continued regulatory compliance throughout the device lifecycle.

Read More »

IVDR Scientific Validity Explained: A Complete Guide for Manufacturers

Scientific Validity is the first pillar of IVDR Performance Evaluation and provides the scientific foundation demonstrating that an analyte or biomarker is associated with a specific clinical condition or physiological state. This guide explains Scientific Validity under Regulation (EU) 2017/746, including literature reviews, Scientific Validity Reports, Annex XIII requirements, evidence appraisal and how Scientific Validity supports successful CE marking.

Read More »

IVDR Performance Evaluation Explained: A Complete Guide for Manufacturers

Performance Evaluation is one of the most important requirements under the EU In Vitro Diagnostic Regulation (IVDR). Every manufacturer must demonstrate that their in vitro diagnostic medical device achieves its intended purpose through robust scientific validity, analytical performance and clinical performance evidence. This guide explains every stage of IVDR Performance Evaluation, including Performance Evaluation Plans (PEPs), Performance Evaluation Reports (PERs), Post-Market Performance Follow-up (PMPF) and how Performance Evaluation supports successful CE marking under Regulation (EU) 2017/746.

Read More »
patient guard
Patient Guard

Sign up to our newsletter

Be the first to hear industry news and how Patient Guard can help you.

Get the latest updates on medical device regulation

Sign up to our newsletter and we’ll deliver news and insights straight to your inbox.

Get the Medical Device Technical Checklist

Thank you! The checklist is now ready to download.