Medical Device Design and Development

Planning for the design and development of a medical device is a requirement of regulatory systems. All manufacturers of Medical Devices are required to maintain a Quality Management System, in the USA this is determined under the CFR Title 21 part 820. Within the European Union, it is placed within demonstrating certificated compliance with international standard EN ISO 13485. The USA has recently indicated that they will also use the ISO 13485 standard for the requirements of manufacturers being compliant for QMS purposes. 
Medical Device Life Cycle following the Plan, Do, Check, Action model. Used by patient guard to explain medical device design and development life cycle which start with medical device regulations, then standards, then design inputs, then risk management, then design outputs, prototypes, design verification and validation, manufacturing specifications and finally design history, before the cycle starts again at regulations.
Facebook
X
LinkedIn

The Medical Device Design and Development Process

The medical device design and development process is a critical process that ensures
regulatory compliance and product safety. In this guide, we’ll share the lifecycle of medical device design
and development, highlighting relevant regulations, standards and documentation for product
manufacturers.

A man with a spyglass looking at data n for form of graphs on a screen. This image is used by patient guard to discuss their content relating to the design and development of medical devices and IVDs.

The Medical Device Design and Development Lifecycle

The lifecycle of the design and development of
medical devices and IVDs is meticulously
documented to ensure compliance and
traceability. In the following section, we’ll outline
the stages and documentation needed to meet
regulatory requirements.

Medical Device Life Cycle following the Plan, Do, Check, Action model. Used by patient guard to explain medical device design and development life cycle which start with medical device regulations, then standards, then design inputs, then risk management, then design outputs, prototypes, design verification and validation, manufacturing specifications and finally design history, before the cycle starts again at regulations.

Medical Device and IVD Regulation

Regulations define the risk, safety, and performance requirements that must be met by manufacturers.
There are numerous different standards, which apply to different regions, groups and specific medical
devices. So, when you start to design and develop a medical device or IVD, it’s important to review the
regulations specific to its intended use. With an understanding of these, you can discuss them with the
design team, before they start the design and development process.

EU MDR and IVDR

In the European Union, the relevant regulations are:

These regulations outline the General Safety and Performance Requirements (GSPRs) in Annex I. These need to be assessed against the concept and intended use of the device being designed and developed:

UK Medical Device Regulations 2002

In the UK, compliance with the Medical Device Regulations 2002 is essential. Relevant regulations, which
outline the Essential Requirements (ERs) for design and development are:

The ERs can be found in Annex 1 of each of these directives, and should be assessed against the concept
and intended use of the medical device being developed.

International Medical Device Standards

Harmonised standards ensure that medical devices meet international safety and performance
benchmarks. Here are some important medical device-related standards that are accepted
internationally.

All medical devices:

  • ISO 14971 – Medical devices. Application of risk management to medical devices
  • ISO 20417 – Medical devices. Information to be supplied by the manufacturer
  • ISO 15223 – Medical devices. Symbols to be used with information to be supplied by the manufacturer – General requirements
  • ISO 13485 – Medical devices. Quality management systems. Requirements for regulatory purposes
  • ISO 62366 – Medical devices – Application of usability engineering to medical devices

Medical devices containing software or software as a standalone medical device:

  • IEC 62304 – Medical device software. Software life-cycle processes

Electrical devices:

  • ISO 60601-1 series – General requirements for safety – Collateral standard. Safety requirements
    for medical electrical systems
  • ISO 60601-2 series – Particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnosis

Biological Evaluation for patient contacting devices:

  • ISO 10993 series – Evaluation and testing within a risk management process

Sterilised medical devices:

  • ISO 11135 – Sterilization of health care products. Ethylene oxide – Requirements for
    development, validation and routine control of a sterilization process for medical devices
  • ISO 11137 – Sterilization of health care products – Radiation – Requirements for development,
    validation and routine control of a sterilization process for medical device
  • ISO 17665 – Sterilization of health care products – Moist heat – Requirements for the
    development, validation and routine control of a sterilization process for medical devices
  • ISO 11607-1 – Packaging for terminally sterilized medical devices – Requirements for materials,
    sterile barrier systems and packaging systems
  • ISO 11607-2 – Packaging for terminally sterilized medical devices – Validation requirements for
    forming, sealing and assembly processes

For Specific Equipment:

  • ISO 60601-2-24 – Medical electrical equipment – Particular requirements for the basic safety
    and essential performance of infusion pumps and controllers
    ISO 3826-4 – Plastics collapsible containers for human blood and blood components –
    Aphaeresis blood bag systems with integrated features

Design and Development Plan

Once you have a good understanding of the regulatory landscape, it is mandatory to develop a
comprehensive design and development plan. This should cover the stages of design, reviews,
verification, validation, and design transfer activities. It also includes responsibilities, authorities, and
traceability methods.

Image of a computer screen with a tick against some test and a cross against other text. This image is used by patient guard to describe medical device and IVD design and development inputs and outputs.

Quality Management Systems (QMS)

Before starting the medical device design and development process, it is important to commit to and maintain a Quality Management System (QMS) as laid
out in the following standards:

The Design and Development process is set out within these QMS standards. And planning should cover
the following as a minimum.
https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820/subpart-C/section-820.30

  • Design and development stages
  • Review(s) at each design and development stage
  • Verification, validation, and design transfer activities appropriate to each design and development stage
  • Responsibilities and authorities for design and development
  • Methods ensuring traceability of design and development outputs to design and development inputs
  • Necessary resources, including required experience / qualifications of personnel

Design and Development Stages

Designing medical devices involves several critical steps, each defined by EN ISO 13485 standards.

  • Design Inputs include functional, performance, usability, and safety requirements, along with
    relevant regulatory requirements and risk management outputs.
  • Design Outputs must meet input requirements, provide necessary information for purchasing, production and service provision, and specify product characteristics essential for safe use.

Design and Development Reviews

Systematic reviews ensure that design outputs meet requirements. These involve representatives from relevant functions and identify necessary actions required.

Design and Development Verification

Verification makes sure that design outputs align with design input requirements. Plans should include methods, acceptance criteria, and rationalised statistical techniques. If the medical device is intended to connect to another medical device, verification should also confirm it meets design inputs when connected.

Design Validation

Conducted on a representative product i.e., from initial production units, batches or equivalent, validation confirms that a product meets the requirements for its intended use.
It should also include methods, acceptance criteria, and rationalised statistical
techniques. This involves clinical evaluations or performance evaluations and is completed before product release.

Design and Development Transfer

Documented procedures ensure that design outputs are suitable for manufacturing. These procedures verify that production can meet product requirements

Control of Design and Development Changes

Changes in design must be controlled, reviewed, verified, validated, and approved to ensure they meet regulatory requirements and maintain product integrity

Risk Management in Design and Development

Risk management is crucial throughout the medical device design and development process. In accordance with ISO 14971, risk management activities should be planned, executed, and documented throughout the product lifecycle.

Risk Management Plan

As a minimum the risk management plan should consider the following:
  • Scope of planned risk management activities
  • Assignment of responsibilities and authorities
  • Requirements for reviewing risk management activity
  • Criteria for risk acceptability
  • Methods and criteria for evaluating overall residual risk
  • Activities for verifying risk control measures
  • Activities for reviewing production and post-production information

Risk Identification and Evaluation

Risks are identified at the start of the design and development process. They are then analysed and scored based on their likelihood and severity. Mitigation measures are then applied to reduce risks to acceptable levels.

Table Medical Devices

Risk Control and Evaluation

Mitigated risks are re-evaluated to make sure they are within acceptable limits. A risk-benefit analysis then determines if the benefits of the device outweigh potential risks.

Mitigating Risk: An Example

Mitigation controls can include the use of standards. For example, a device which has patient contact and is sterilised using ethylene oxide.

Ethylene Oxide is known to cause irritation, organ damage, mutagenicity and carcinogenicity in human and animals, and reproductive effects in animals. So, the risk is ethylene oxide residuals on the device post-sterilisation and packaging, which could come into contact with the user.

Risk control measures would be to ensure that the ethylene oxide sterilisation process has been followed (ISO 11135) and testing carried out on the final finished post-sterilised device (ISO 10993-7).

Ethylene oxide sterilization residuals, and the test results evaluated as part of the Biological Evaluation Report by an expert toxicologist in line with ISO 10993-1 – Evaluation and testing within a risk management process.
The risk should then be scored again based on the control measure.

Proof of Concept

Proof of concept testing demonstrates that critical design features function as intended. This early-stage testing is crucial for identifying potential issues before clinical evaluations.

Design History

A comprehensive design history is maintained within the Quality Management System (QMS). This documentation includes detailed records of the design and development process.

By following these guidelines, medical device manufacturers can ensure that products meet regulatory requirements and are safe for use. This meticulous documentation and adherence to standards throughout design and development are critical for successful product  development.

How can Patient Guard Help?

Patient Guard has helped hundreds of medical device and IVD manufacturers since 2017 with their medical device design and development. If you are designing and developing a medical device we can assist you with the regulatory requirements and technical documentation you need to provide as part of regulatory requirements. Contact us to see how we can assist with your medical device project. 

FAQs

The design and development process for medical devices involves transforming an initial concept into a market-ready product. It includes defining user needs, creating a design plan, prototyping, testing, and ensuring regulatory compliance.

Why it matters: A structured process ensures the device is safe, effective, and meets regulatory requirements such as the EU MDR, FDA regulations, or ISO 13485.

The process typically includes:

  1. Concept and Feasibility: Defining user needs and technical feasibility.
  2. Design Planning: Developing a structured plan, including regulatory considerations.
  3. Design Inputs: Documenting specifications based on user needs and safety requirements.
  4. Design Outputs: Creating detailed designs, including prototypes and schematics.
  5. Design Verification and Validation: Ensuring the design meets specifications and performs as intended.
  6. Transfer to Manufacturing: Scaling production with quality controls in place.

Pro tip: Aligning the process with regulatory standards like ISO 13485 ensures smoother market entry.

Risk management, as outlined in ISO 14971, is integral to medical device design. It involves:

  • Identifying Risks: Assessing potential hazards associated with the device.
  • Mitigating Risks: Designing controls to minimize risks.
  • Residual Risk Evaluation: Ensuring remaining risks are acceptable.
  • Ongoing Monitoring: Updating risk assessments based on post-market data.

Key takeaway: Incorporating risk management early reduces safety issues and ensures compliance.

  • Design Verification: Confirms the design meets input specifications (e.g., performance, safety).
  • Design Validation: Ensures the device meets user needs and intended purposes under real-world conditions.

Why it matters: Verification and validation are regulatory requirements under frameworks like FDA QSR and EU MDR, and they ensure the device functions safely and effectively.

Regulations like the EU MDR, IVDR, and FDA guidelines shape every stage of design and development. Key requirements include:

  • Defining intended use and classification early in the process.
  • Ensuring a Quality Management System (QMS) aligned with ISO 13485.
  • Documenting design controls, risk management, and verification/validation results.
  • Preparing for regulatory submissions with complete technical documentation.

Key insight: Integrating compliance from the start avoids costly redesigns or approval delays.

Yes! Patient Guard offers expert support throughout the design and development process, including:

  • Developing regulatory-compliant design and development plans.
  • Assisting with risk management and design verification/validation.
  • Preparing technical documentation for CE marking, FDA, or UKCA submissions.
  • Guiding compliance with standards like ISO 13485 and ISO 14971.

Why choose Patient Guard: With extensive experience in regulatory frameworks and over 500 successful projects, we ensure your device is designed for safety, performance, and compliance.

Resources

Facebook
X
LinkedIn

Most Popular

Medical Device Design and Development

Planning for the design and development of a medical device is a requirement of regulatory systems. All manufacturers of Medical Devices are required to maintain a Quality Management System, in the USA this is determined under the CFR Title 21 part 820. Within the European Union, it is placed within demonstrating certificated compliance with international standard EN ISO 13485. The USA has recently indicated that they will also use the ISO 13485 standard for the requirements of manufacturers being compliant for QMS purposes. 

Read More »

Post Market Surveillance – Medical Devices

Post-market surveillance (PMS) is a pivotal aspect that governs the continued monitoring and assessment of medical devices after they have entered the market. Across the EU, UK, and USA, distinct regulatory frameworks shape PMS protocols, fostering safety, and innovation simultaneously.

Read More »

December 2024 News Letter

Welcome to our December 2024 news letter, here we share the latest medical device regulatory news from the EU, UK and the USA as well as updated or new medical device standards that have been issued.

Read More »
patient guard
Patient Guard

Sign up to our newsletter

Be the first to hear industry news and how Patient Guard can help you.

Get the Medical Device Technical Checklist

Thank you! The checklist is now ready to download.

Get the Medical Device Technical Checklist

Complete the form below and receive instant access.

Speak to one of our medical device consultants

For help with the checklist or other aspects of your compliance journey, please reach out to us at Patient Guard and our experts would be happy to help.

UK Office

Get the latest updates on medical device regulation

Sign up to our newsletter and we’ll deliver news and insights straight to your inbox.

Patient Guard Regulatory Affairs and Quality Assurance

Do you need support with Medical Device or IVD compliance?

We can help you!